It is quite easy to get surface scans with large scan ranges. However,
achieving atomic scale is more demanding. Here are some hints that should help you:
- It is essential to apply short scanning times to minimize thermal drift effects (set time/line to a
minimum value).
- The Rotation parameter needs to be optimized. Take some images at different values and use
the one with the best resolution.
- Bright light and movement near the STM setup interfere with the measurement.
Avoid abrupt movements and walking around in the room.
- If a measurement looks good, wait a few minutes to obtain stable images.